Issue: Air tank pressure switch turns ON 15 times in 1 hours.
Probable Cause:
- Leak on nitrogen generator air tank. Check Valve sticking open could cause the leak.
- Excess air bleed regulator closed.
- Pressure switch needs to be adjusted or replaced.
- Pre-filters need to be replaced.
- Membrane inlet solenoid is stuck open due to debris.